Codman Hakim Shunt System

The Codman Hakim shunt valve is manufactured by Codman Inc. and can be set to 18 different pressure levels ranging from 30 to 200 mmH2O. General information and illustrations for reading the pressure level using the radiographic indicator can be found in the official manufacturer information.

Tabular overview

Adjustable opening pressure

MRI compatibility

Yes, from 30 to 200 mm H2O

MR conditional. Shunt valve settings must be checked after MRI[^1]

MRI compatibility

The Codman Hakim shunt valve can be used in MRI systems up to 3 Tesla in accordance with the official manufacturer information without impairing valve function. However, the manufacturer explicitly recommends checking the shunt valve setting after MRI because it may be altered by the magnetic fields in the MRI system[^1]. According to the official manufacturer information, there is no degradation in the valve's adjustability at 1.5-Tesla MRI. At 3-Tesla MRI, reduced adjustability of the shunt valve may occur after 10 MRI examinations.

Reading the pressure level

The currently set pressure level of Codman Hakim shunt valves can be reliably determined by obtaining an X-ray of the shunt valve.

Codman Hakim shunt valve pressure levels
Visualization of the various valve pressure levels of a Codman Hakim shunt valve on X-ray.

Adjusting the pressure level

Codman Hakim shunt valves are adjusted using a specialized device provided by Codman. According to the manufacturer information, it is recommended not to change the opening pressure by more than 40 mmH20 within 24 hours.

Imaging

Codman Hakim shunt valve on a lateral X-ray
X-ray of the Codman Hakim shunt valve on a lateral radiograph.

Further information

Editorial transparency

Authorship & revisions

Who created and updated this article, and which changes are documented.

Article created by
Last edited:
on
Revision history
3 edits · 1 contributor
Author list:
Show revision history 3 entries

3 edits by 1 author

Edit · Alaric Steinmetz ·

No content change recorded

Edit · Alaric Steinmetz · (+13 −13)
  • Text changed: “The Codman Hakim shunt valve is produced by Codman Inc. and can be adjusted to 18 different pressure settings ranging from 30 to 200 mmH2O. The official manufacturer information contains general information and illustrations for reading the pressure setting using the radiographic indicator.”“The Codman Hakim shunt valve is manufactured by Codman Inc. and can be set to 18 different pressure levels ranging from 30 to 200 mmH2O. General information and illustrations for reading the pressure level using the radiographic indicator can be found in the official manufacturer information.”
  • Text changed: “According to the official manufacturer information , the Codman Hakim shunt valve can be used in MRI systems up to 3 Tesla without impairing valve function. However, the manufacturer explicitly recommends checking the shunt valve setting after MRI, as it may be altered by the magnetic fields during MRI[^1]. According to the official manufacturer information, there is no degradation in the adjustability of the valve at 1.5-Tesla MRI. At 3-Tesla MRI, reduced adjustability of the shunt valve may occur after 10 MRI examinations.”“The Codman Hakim shunt valve can be used in MRI systems up to 3 Tesla in accordance with the official manufacturer information without impairing valve function. However, the manufacturer explicitly recommends checking the shunt valve setting after MRI because it may be altered by the magnetic fields in the MRI system[^1]. According to the official manufacturer information, there is no degradation in the valve's adjustability at 1.5-Tesla MRI. At 3-Tesla MRI, reduced adjustability of the shunt valve may occur after 10 MRI examinations.”
  • Text changed: “Reading the pressure setting”“Reading the pressure level”
  • Text changed: “The currently adjusted pressure setting of Codman Hakim shunt valves can be reliably read using a radiograph of the shunt valve.”“The currently set pressure level of Codman Hakim shunt valves can be reliably determined by obtaining an X-ray of the shunt valve.”
  • Text changed: “Adjusting the pressure setting”“Adjusting the pressure level”
  • Text changed: “Codman Hakim shunt valves are adjusted using a special device provided by Codman. According to the manufacturer information, it is recommended that the opening pressure not be changed by more than 40 mmH20 within 24 hours.”“Codman Hakim shunt valves are adjusted using a specialized device provided by Codman. According to the manufacturer information, it is recommended not to change the opening pressure by more than 40 mmH20 within 24 hours.”
  • Text changed: “Additional information”“Further information”

Article created on · Alaric Steinmetz

Image viewer

Image

Open original